Fume Cupboards Service & Validation
Our thorough service includes the following:
- General conditional checks
- Pre-filter conditional check
- Workstation functional checks
- Airflow readings in accordance with BS EN 14175 (2006)
- Filter integrity tests on any HEPA filters BS EN 14644-3 (2005)
- Airflow visualisation for operator protection
Laminar Flow Cabinets Service & Validation
Our thorough service includes the following:
- General conditional checks
- Pre-filter conditional check
- Workstation functional checks
- Airflow readings in accordance with BS EN 14644-3
- HEPA filter integrity tests BS EN 14644-3 (2005)
- Airborne particle counts in accordance with BS EN 14644-1 (1999) and EU GMP Annex 1 (February 2008)
- Airflow visualisation (on customer request)
Microbiological Safety Cabinets (MSC) Service & Validation
Our thorough service includes the following:
- General conditional checks
- Pre-filter conditional check
- Workstation functional checks
- Airflow readings in accordance with BS EN 12469 (2000)
- HEPA filter integrity tests BS EN 14644-3 (2005)
- Airborne particle counts in accordance with BS EN 14644-1 (1999) and EU GMP Annex 1 (February 2008)
- KI discuss operator protection test (optional test)
Pharmaceutical Isolator Service & Validation
Our thorough service includes the following:
- General conditional checks
- Pre-filter conditional check
- Workstation functional checks
- Airflow readings in accordance with BS EN 14644-3 (2005)
- Pressure differentials are recorded in accordance with BS EN 14644-3 (2005)
- HEPA filter integrity tests BS EN 14644-3 (2005)
- Airborne particle counts in accordance with BS EN 14644-1 (1999) and EU GMP Annex 1 (February 2008)
- Breach velocity tests will be carried out in accordance with BS EN 14644-7 (1999)
- Pressure decay tests will be carried out in accordance with BS EN 14644-7 (1999) (optional test)
Powder Control Booth Service & Validation
Our thorough service includes the following:
- General conditional checks
- Pre-filter conditional check
- Workstation functional checks
- Airflow readings in accordance with BS EN 14644-3 (2005)
- HEPA filter integrity tests BS EN 14644-3 (2005)
- Airborne particle counts in accordance with BS EN 14644-1 (1999) and EU GMP Annex 1 (February 2008) (optional test)
PCR Workstation & Powder Control Booth Service & Validation
Our thorough service includes the following:
- General conditional checks
- Pre-filter conditional check
- Workstation functional checks
- Airflow readings in accordance with BS EN 14644-3 (2005)
- HEPA filter integrity tests BS EN 14644-3 (2005)
- Airborne particle counts in accordance with BS EN 14644-1
(1999) and EU GMP Annex 1 (February 2008) (optional test)